Early Alert: Outset Medical Updates Use Instructions for the Tablo Console and the TabloCart with Prefiltration Drawer

SEE FULL FDA NOTICE

The FDA is aware that Outset Medical has issued a letter to affected customers recommending all Tablo hemodialysis systems have updated use instructions prior to continued use. Affected devices: 

  • Power cord with clear ends (PN-0005746) for the Tablo Console and TabloCart with Prefiltration Drawer 
  • Power cord with gray ends (PN-0008460) for the Tablo Console and TabloCart with Prefiltration Drawer 

Outset Medical has received reports of overheating at the connection between the power cord and the power inlet on the Tablo Console and on the TabloCart with Prefiltration Drawer. Outset Medical identified that these events occur when the power cord connector is not held firmly in the power inlet resulting in movement of the connector, increased electrical resistance, and heat generation at the connection. Reports have involved both the original power cord with clear ends (PN-0005746) and the current power cord with gray ends (PN-0008460). 

If overheating occurs at the connection, a person touching the overheated connector could receive a first- or second-degree burn, and the Tablo Console may lose power, which could delay or interrupt a treatment or cause a treatment to be missed.  

As of November 10, 2026, Outset Medical has reported no serious injuries or deaths associated with this issue.