SEE FULL FDA NOTICE
The FDA is aware that B. Braun has issued a letter to affected customers recommending certain Hemodialysis Bloodlines be removed from where they are used or sold. Affected devices:
- Product Description: Streamline Bloodline Set for Dialog+
- Model: SL-2010M2096
- UDI-DI: 04046964367786; 04046964367779
- Impacted Lots: A2600099, A2600128, A2600136
B. Braun stated that the required solvent application step during assembly was not performed, resulting in inadequate bonding of the dialyzer connector. As a result, the bloodline connector may detach or leak during priming or dialysis treatment, potentially leading to blood loss, clotting and occlusion of the extracorporeal circuit, and/or air ingress into the blood circuit.
As of October 10, 2026, B. Braun has reported no serious injuries or deaths associated with this issue.