Early Alert: B. Braun Removes Streamline Bloodline Set for Dialog+

SEE FULL FDA NOTICE

The FDA is aware that B. Braun has issued a letter to affected customers recommending certain Hemodialysis Bloodlines be removed from where they are used or sold. Affected devices:  

  • Product Description: Streamline Bloodline Set for Dialog+  
  • Model: SL-2010M2096 
  • UDI-DI: 04046964367786; 04046964367779 
  • Impacted Lots: A2600099, A2600128, A2600136 

B. Braun stated that the required solvent application step during assembly was not performed, resulting in inadequate bonding of the dialyzer connector. As a result, the bloodline connector may detach or leak during priming or dialysis treatment, potentially leading to blood loss, clotting and occlusion of the extracorporeal circuit, and/or air ingress into the blood circuit. 

As of October 10, 2026, B. Braun has reported no serious injuries or deaths associated with this issue.